See it for yourself: raise or lower the sample
The diagram shows the sample height, drawn many times larger than scale. The corundum pattern under the slider shows what the detector records: dashed is where the peaks belong, solid is where they land.
| Reflection | Theoretical 2θ | Collected 2θ | Shift | Apparent d error |
|---|
Simulated scan of the NIST SRM 1976c sintered corundum (α-Al₂O₃) plate, Cu Kα₁/Kα₂. Peak positions use the certified lattice parameters (a = 0.4759092 nm, c = 1.299337 nm) and match the certificate's Table 3; positions in the table are Kα₁. Relative intensities are the certified values from Table 1, which reflect the plate's texture (NIST SRM 1976c Certificate of Analysis). Reflections marked * aren't certified and are estimates. In the diagram, lowering the surface by s moves the point where the beam strikes it farther along the incident beam, so the same reflection reaches the goniometer circle at a lower 2θ; the two diffracted beams are offset by 2s cos θ, which is where Equation 1 in the article comes from. The diagram is drawn for a peak at 70° 2θ. Shifts follow that equation, with positive s meaning the sample surface sits below the reference plane. Only displacement is modeled here; every other error is set to zero.